S/T Mode
Setting the rise time
Rise time is the speed with which inspiratory pressure rises to the set (target) pressure
- Set to fastest rise time tolerated
- 1 is the fastest rise, 5 the slowest
- Too slow a rise time may exacerbate dyspnea in acute RF patients as this may deprive patient of needed flow
- Too fast a rise time may cause discomfort and increased leaks15
- If the rise time is insufficient to reach the desired inspiratory time, decrease the rise time setting (eg: change from 5 to 3, therefore resulting in a faster rise time)
- As the patient’s condition stabilizes the rise time setting should be increased (eg: on the V60 a change from 1 to 3, therefore resulting in a slower rise time)