Respironics
Whisper Swivel
- Intentional leak
- Fixed orifice variable flow. Note: A minimum EPAP of 4 cmH20 is supplied with V60 to ensure sufficient CO2 flushing in single limb circuit
- Multi patient use exhalation port that provides a continuous leak path
- A proximal line port must be connected to the mask or added connector for use with the V60
Disposable Exhalation
Port (DEP)
- Intentional leak
- Single use, fixed orifice variable flow
Note: A minimum EPAP of 4 cmH20 is supplied with V60 to ensure sufficient CO2 flushing in single limb circuit.
ComparePlateau Exhalation Valve
(PEV)11
- Intentional leak
- Reusable exhalation port for use with integrated proximal pressure port
- Variable orifice with fixed flow which provides higher leak than fixed orifices (DEP or Whisper Swivel) at lower expiratory pressures that will aid in flushing CO2 if there is a concern for rebreathing
Comparison of Exhalation Port performance
Leak Flow Rate vs Circuit Pressure of the
Disposable Exhalation Port, Whisper Swivel,
and PEV.
The PEV leak rate is higher than fixed orifices
(DEP or Whisper Swivel) at lower expiratory
pressures, which will aid in flushing CO2 if
there is a concern for rebreathing.