S/T Mode

Setting the rate

  • Rate in S/T mode creates a time window in which the patient must trigger

Example

if backup rate is set at 12 bpm, the ventilator will look for a spontaneous breath within 5 seconds, if none is detected a machine triggered PS breath will be initiated.

Rate

Since V60 is intended to augment ventilation in patients whom are spontaneously breathing, the rate should be set:

  • as a backup in case of apnea
  • slightly below the expected normal rate for this patient. This will allow initial drop in spontaneous breath rate (expected with reduced WOB) while avoiding a machine triggered breath

Example

Patient spontaneous breath rate upon NIV initiation = 30 bpm

Normal resting rate expected 18 bpm
Consider setting back up rate 10-12 bpm or lower if prolonged expiratory times are needed in the case of severe COPD exacerbations.

Consider setting low rate alarm at 14 bpm