S/T mode

The S/T (spontaneous/timed) mode guarantees breath delivery at the user-set rate. It delivers pressure-controlled, time-cycled mandatory and pressure- supported spontaneous breaths, all at the IPAP pressure level.

If the patient fails to trigger a breath within the interval of time determined by the Rate setting, the ventilator triggers a mandatory breath with the set I-Time. You set no patient triggering or cycling sensitivities: the patient triggers and cycles based on the ventilator’s Auto-Trak Sensitivity algorithms*.

Mode selection: Tap button on screen for more information on each mode type

While Auto-Trak is effective on most patients, an optional upgrade allows the titration of trigger and cycle sensitivities within the Auto-Trak algorithm. This option is called Auto-Trak+