S/T Mode

Adjust EPAP setting

EPAP = PEEP

  • If PS level needs to be maintained, a change in EPAP should be followed by a similar change in IPAP setting.
  • Use the arrows to increase or decrease EPAP

Initiate patient on EPAP on 4-6 cmH20. Ensure resulting PS is adequate.

Guidance on increasing EPAP

  • Signs of air trapping
  • Wasted efforts on flow waveform
  • Observed delay between patient effort and vent triggering due to AutoPEEP
  • Low PaO2 or SpO2

Guidance on decreasing EPAP

  • Cardiovascular comprise in non-cardiogenic pulmonary patients
  • Discomfort
  • Active exhalation (active expiratory muscle following resolution of acute phase)

Any increase or decrease of EPAP will affect the level of pressure support delivered