Assessment of
patient-ventilator synchrony
Ensure the patient’s efforts are in synchrony with the ventilator’s response.
Place hand on chest and watch graphics on screen
- Ensure the device switches to inspiration phase with each of the patient's inspiratory efforts. This can be done by watching the breath type indicator bar switches to Turquoise at the same time.
- Determine if patient’s inspiration is in time with onset of IPAP.
- Ensure that patient flow demand is met.
- Confirm that end of patient’s inspiration matches with transition to EPAP. This can be done by ensuring the breath type indicator bar switches to dark blue.
NIV hints and tips
- Coaching the NIV patient is critical
- Titrate patients aggressively, initial parameters may help
many patients, but in others, failure to adequately titrate may
result in NIV failure - Remember dedicated NIV Devices are not PEEP compensated;
therefore both IPAP and EPAP most be adjusted to maintain desired
Pressure Support. IPAP-EPAP=PS IPAP ≠ PS - Keep inspiratory pressures < 20 cmH20 to avoid gastric distension
- Use humidification to prevent drying of secretions
and mucous membranes - Continue to assess patient’s comfort
- Mask leaks should be monitored and kept between 7 - 25 lpm.
Too tight = pressure sores. Too loose = drying, conjunctivitis, FiO2
control loss. - Consider a mask rotation strategy to optimize patient comfort
and tolerance of NIV.