Optimize synchrony
| Symptom/sign | Likely cause | Correction |
|---|---|---|
| Ventilator triggering while patient is breathing spontaneously |
Backup rate is set higher than patient rate |
Lower backup rate to 10-12 or below patient's spontaneous rate |
| Trigger delay or missed trigger | AutoPEEP Patient leak > 60 lpm with normal intentional leak (Mask leak 1 & DEP) or (total leak › 80-90 lpm Patient too weak to trigger |
Increase EPAP (without corresponding increase in IPAP to reduce pressure support.) Reduce patient leak by readjusting or changing mask If patient leak is between 7-25 lpm and intentional leak is high (a DEP is used with a Leak 2 or greater mask) - remove DEP or alternatively change to leak 1 elbow. However, always ensure there is sufficient intentional leak to flush CO2. Reassess whether patient is still a candidate for NIV Reduce sedation Consider increasing trigger sensitivity with Auto-Trak + (optional feature) |
| Double notching on inspiratory pressure waveform |
Insufficient inspiratory flow | Decrease the rise time setting (eg: from 3 to 1) to make initial rise to pressure faster thereby provide more flow. Switch to ventilator with high flow capacity such as the V60 at 240lpm. |
| Premature cycling | Muscle weakness | Reassess whether patient is still a candidate for NIV Reduce sedation Consider decreasing cycle sensitivity with AT+ (optional feature) |
| Delayed cycling | I-time is too long in machine triggered breath |
Shorten I-time |