Optimize synchrony

Symptom/sign Likely cause Correction
Ventilator triggering while patient
is breathing spontaneously
Backup rate is set higher
than patient rate
Lower backup rate to 10-12 or below patient's spontaneous rate
Trigger delay or missed trigger AutoPEEP


Patient leak > 60 lpm with
normal intentional leak
(Mask leak 1 & DEP) or 
(total leak › 80-90 lpm


Patient too weak to trigger
Increase EPAP (without corresponding increase in IPAP to reduce
pressure support.)

Reduce patient leak by readjusting or changing mask
If patient leak is between 7-25 lpm and intentional leak
is high (a DEP is used with a Leak 2 or greater mask) - 
remove DEP or alternatively change to leak 1 elbow. However,
always ensure there is sufficient intentional leak to flush CO2.

Reassess whether patient is still a candidate for NIV
Reduce sedation
Consider increasing trigger sensitivity with Auto-Trak + 
(optional feature)
Double notching on inspiratory
pressure waveform
Insufficient inspiratory flow Decrease the rise time setting (eg: from 3 to 1) to make initial rise
to pressure faster thereby provide more flow.
Switch to ventilator with high flow capacity such as the V60 at 240lpm.
Premature cycling Muscle weakness Reassess whether patient is still a candidate for NIV
Reduce sedation
Consider decreasing cycle sensitivity with AT+ (optional feature)
Delayed cycling I-time is too long in
machine triggered breath
Shorten I-time