“While there is not clear consensus on whether or not additional heat and humidity are always necessary when the upper airway is not by-passed, such as in non-invasive mechanical ventilation (NIV), active humidification is highly suggested to improve comfort.”

HME recommendations
Humidification of inspired gas during mechanical ventilation is mandatory when an endotracheal or tracheostomy tube is present, but optional with NIV.

Use of an HME is contraindicated in patients on NIV with large mask leaks, as the patient does not exhale enough tidal volume to replenish heat and moisture to adequently condition the inspired gas.

Active humidification is suggested for NIV, as it may improve adherence and comfort (2B).

Passive humification is not
recommended for NIV (2C).